Characterization of single-sided charge-sharing CZT strip detectors for gamma-ray astronomy


Creative Commons License

Donmez B., MACRI J. R., RYAN J. M., LEGERE J. S., MCCONNELL M. L., WIDHOLM M., ...Daha Fazla

Conference on Hard X-Ray and Gamma-Ray Detector Physics and Penetrating Radiation Systems VIII, California, Amerika Birleşik Devletleri, 14 - 17 Ağustos 2006, cilt.6319 identifier identifier

  • Yayın Türü: Bildiri / Tam Metin Bildiri
  • Cilt numarası: 6319
  • Doi Numarası: 10.1117/12.683695
  • Basıldığı Şehir: California
  • Basıldığı Ülke: Amerika Birleşik Devletleri
  • Anahtar Kelimeler: CZT, strip detectors, gamma-ray
  • Akdeniz Üniversitesi Adresli: Hayır

Özet

We report progress in the study of thick single-sided charge-sharing cadmium zinc telluride (CZT) strip detector modules designed to perform spectroscopy and 3-D imaging of gamma-rays. We report laboratory measurements including spectroscopy, efficiency and 3-D imaging capability of prototype detectors (15 x 15 x 7.5mm(3)) with 11 x 11 unit cells. We also report on Monte Carlo simulations (GEANT4v7.1) to investigate the effect of multihits on detector performance in both spectroscopy and imaging. We compare simulation results with data obtained from laboratory measurements and discuss the implications for future strip detector designs.