First successful percutaneous treatment of a totally occluded HeartWare outflow graft: Case report and literature review.


Creative Commons License

Kemaloğlu C., Altekin R. E., Bayezid Ö.

Anatolian journal of cardiology, cilt.19, ss.341-345, 2018 (SCI-Expanded) identifier identifier identifier

  • Yayın Türü: Makale / Tam Makale
  • Cilt numarası: 19
  • Basım Tarihi: 2018
  • Doi Numarası: 10.14744/anatoljcardiol.2018.76736
  • Dergi Adı: Anatolian journal of cardiology
  • Derginin Tarandığı İndeksler: Science Citation Index Expanded (SCI-EXPANDED), Scopus, TR DİZİN (ULAKBİM)
  • Sayfa Sayıları: ss.341-345
  • Anahtar Kelimeler: heart failure, ventricular assist device, outflow graft, balloon angioplasty, VENTRICULAR ASSIST DEVICE, THROMBOSIS, STENOSIS
  • Akdeniz Üniversitesi Adresli: Evet

Özet

As the number of implanted left ventricular assist devices (LVADs) used increases, the frequency of chronic complications encountered also increases. The pause of blood flow in the outflow graft is a rare, but fatal, complication. The aim of this article was to present the case of a patient in whom HeartWare outflow graft occlusion was removed by balloon angioplasty and to examine the treatment modalities of HeartWare outflow graft occlusions that have been percutaneously performed to date. The literature was searched for percutaneous interventions on outflow grafts of the left ventricular assist devices. The results of six patients who underwent interventions on outflow grafts were analyzed. Three of six patients with HeartWare outflow graft stenosis were treated with covered stents, while the remaining three were treated with bare metal stents. All procedures were applied successfully. Percutaneous interventions can be performed with appropriate equipment in patients with HeartWare outflow graft stenosis or total occlusion.