Evaluation of non-ionizing radiation around the shortwave diathermy devices


Ozen U., Comlekci S., Cerezci O., Onural A.

IEEE International Symposium on Electromagnetic Compatibility (EMC), İstanbul, Turkey, 11 - 16 May 2003, pp.824-827, (Full Text) identifier identifier

  • Publication Type: Conference Paper / Full Text
  • Doi Number: 10.1109/icsmc2.2003.1429034
  • City: İstanbul
  • Country: Turkey
  • Page Numbers: pp.824-827
  • Keywords: short-wave diathermy, EMI, RF radiation
  • Akdeniz University Affiliated: No

Abstract

Short Wave and Microwave equipment are used for treatment purpose in physiotherapy units. Such equipments' stray fields can affect around them in the near field region. This fact has been stressed in current studies. In this way, some "safe operating area"s or safe proximity distances have been recommended. Effects on the health of EM energy, nowadays, have, been discussed topic frequently. So, controlled Medical Aimed Radiofrequency Energy gained more and more importance at the Hospitals. In this study; leakage or stray fields of shortwave diathermy (27.12 MHz) equipments are measured and evaluated in terms of standards and safe proximity or approach distances. Also, placement of various types of treatment equipments is be evaluated in order to avoid over exposure and electromagnetic interference (EMI). A practical measuring methode is for new or old equipment in medical units.